Sumber Aneka Karya Abadi - Your trusted partner for laboratory instrument

Search

HACH ONLINE, 5500sc Silica Analyzer, 1 channel, reagents included

Prod # 5500.KTO.S0.A1U

Product Overview

Spesifikasi :

Accuracy

0–500 µg/L: ±1% or ±1 µg/L of reading, whichever is greater; 500–5000 µg/L: ±5%

Alarm

Four relays output; type: not powered SPDT relays, each rated at
5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded recommended

Certifications

CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 )
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)

Controller Compatibility

sc200, sc1000

Dimensions (H x W x D)

804 mm x 452 mm x 360 mm

Enclosure Rating

IP56 / NEMA 4X

Fitting type

Sample line and sample bypass drain: 6 mm (¼-in.)

Air purge inlet: 6 mm (¼-in.)

Grab Sample

Grab Sample In and Grab Sample Out capability

Light Source

Class 1M LED (light emitting diode) with a peak
wavelength of 810 nm

Lower Limit of Detection (LOD)

0.5 µg/L

Manual Languages

English
French
Spanish
B. Portuguese
Chinese
Japanese
Korean
Thai
German
Italian
C. Portuguese
Czech
Danish
Dutch
Polish
Swedish
Finnish
Bulgarian
Hungarian
Romanian
Lithuanian
Russian
Turkish
Slovak
Slovenian
Croatian
Greek
Estonian

Measuring Principle

Colorimetric

Mounting

Wall, panel or table

Number of Channels

1

1

Operating Humidity

5 to 95 % RH non-condensing (indoor use only)

Operating Temperature Range

5 to 45 °C

Output

4 to 20 mA
4 to 20 mA

Parameter

Silica

Pollution Degree

2/II

Power Requirements (Hz)

50/60 Hz

Power Requirements (Voltage)

100 - 240 VAC

Protection Class

l

Range

0 - 5000 µg/L as SiO2

Reagent Consumption

2 L of each reagent every 90 days with 15 minute cycle time
2 L of each reagent every 90 days with 15 minute cycle time

Relays

Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum

Repeatability

± 0.5 µg/L or ± 1 % of reading, whichever is greater

Response Time

Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature

Sample Flow Rate

55 to 300 mL/min

Sample Pressure

2 to 87 psi (to Preset Pressure Regulator)

Sample Temperature

5 to 50 °C

Storage Conditions

-20 to 60 °C

Warranty

1 year

Weight

44.09 lbs. (20 kg) (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents