Specifications :
Accuracy:Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2:Free Chlorine: ± of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3:Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4:Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection:5 pin, M12 connector
Cable Length:1 m (between gateway and sc controller)
Calibration Method:1-point or 2-point (zero and slope) calibration
Certifications:CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line:1/2 in ID
Connections:1/4 in OD
Controller:Panel only
Controller Compatibility:sc controller platform
Description :CLT10sc Total Chlorine Analyzer (Panel Only) with Combination pH Sensor
Dimensions:mm x 25 mm (7.68 in x 0.98 in)
Sensor Only: 195
Fitting type:USA
Includes:CL10sc Panel, 1 m digital extension cable, Panel Manual, Combination pH Sensor Manual, Chlorine Sensor Manual.
Interferences:Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length:Sensor: 195 mm
Lower Limit of Detection (LOD):30 ppb (0.03 ppm)
Material:Corrosion-resistant materials, fully-submersible
(stainless steel, PVC, silicon rubber and polycarbonate)
Measurement method:Reagentless, electrochemical, three-electrode amperometric system
Mounting:Panel or Wall Mounting
Operating temperature range:0 - 45 °C
Options:Combination pH Sensor
Parameter:Total Chlorine
pH Monitoring Required?:With combination pH sensor
pH Range:4 - 9
Power requirements (Voltage):None
Pressure Range:0.5 bar, no pressure impulses and/or vibrations
Range:0 - 20 ppm Clâ‚‚
Repeatability:30 ppb or 3%, whichever is greater
Response time:Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
Response Time:Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate:30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature:5 - 45°C (41 - 113°F)
Storage conditions:-20 °C to 60 °C dry
Weight:Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?:CL10 sc Panel, 1 m digital extension cable, Panel Manual, Combination pH Sensor Manual, Chlorine Sensor Manual.