TU5300 Turbiditimeter online merupakan pengganti dari turbiditimeter seri 1720E , menggunakan metode ISO 7027-1 dan sempurna untuk filtrasi konvensional dan pengukuran kekeruhan di level rendah. TU5300 menggunakan design optik yang dipatenkan , mengukur lebih banyak sampel dari pada turbiditimeter lainnya, memberikan presisi dan sensitifitas terbaik untuk pengukuran kekeruhan level rendah dengan meminimalisasi variabilitas antar pengukuran. TU5300 Low Rage Laser Turbidimeter dapat menghilangkan nilai ketidakpastian karena menggunakan Detection Tecnology 360º x 90º Detection Tecnology. Dengan kemampuan mengukur hingga 0,002 NTU , TU5300 memberikan standar berikutnya dalam evolusi kekeruhan.
Accuracy : ±2% or 0.01 NTU from 0 - 40 NTU
±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length : 1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration Method : For Formazin and Stablcal®:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range
Custom calibration for up to 6-point calibrations
Certifications : CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking
Communication : None
Controller : Sensor Only
Dimensions (H x W x D) : 249 mm x 268 mm x 190 mm
Enclosure Rating : Electronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting Type : Sample quick connector: ¼-in. for ¼-in. Tubing
Light Source : Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Lower Limit of Detection (LOD) : 0.002 NTU
Material : ASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40,
Thermoplastic Elastomer TPS-SEBS
Model : TU5300sc
Mounting Configurations : Wall Mount
Operating Humidity : Relative humidity: 5 - 95% at different temperatures, non-condensing
Operating Temperature Range : 0 - 50 °C (32 - 122 °F)
Options : Process Head Only
Range : ISO:
0 - 1000 NTU / FNU / TE/F / FTU
0 - 250 EBC
Regulatory : ISO
Repeatability : Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution : 0.0001 NTU / FNU / TE/F / FTU / EBC
Response Time : T90 <45 seconds at 100 mL/min
Sample Flow Rate : 100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure : 6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature : 2 - 60 °C (35 - 140 °F)
Signal Average Time : 30 - 90 seconds
Storage Conditions : -40 - 60 °C (-40 - 140 °F)
Stray Light : <10 mNTU
Units : NTU, FNU, TE/F, FTU, EBC
Verification : Liquid: Stablcal®, Formazin (0.1 - 40 NTU)
Dry: Glass Rod at <0.1 NTU
Weight : 5.95 lbs. (2.7 kg); 11 lbs. (5.0 kg) with all accessories