Sumber Aneka Karya Abadi - Your trusted partner for laboratory instrument

Search

HACH ONLINE, TU5300sc Low Range Laser Turbidimeter with System Check, EPA Version

Prod #LXV445.99.11112

Product Overview

SPECIFICATIONS :

Accuracy:±2% or 0.01 NTU from 0 - 40 NTU
±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length:1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration Method:For Formazin and Stablcal: 
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range

Custom calibration for up to 6-point calibrations
Certifications:CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Australian ACMA Marking
Communication:System Check
Controller:Sensor Only
Controller Compatibility:SC4500, SC1000, SC200
Dimensions (H x W x D):249 mm x 268 mm x 190 mm
Enclosure Rating:Electronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting type:Sample quick connector: ¼-in. for ¼-in. tubing
Instrument:Standard Process Head
Light source:Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Lower Limit of Detection (LOD):0.002 NTU
Material:ASA Luran S 777K / RAL7000, TPE RESIN Elastocon STK40,
Thermoplastic Elastomer TPS-SEBS
Model:TU5300sc
Mounting Configurations:Wall Mount
Operating Humidity:Relative humidity: 5 - 95% at different temperatures, non-condensing
Operating temperature range:0 - 50 °C
Options:Process Head Only
Power requirements (Voltage):None
Range:EPA:
0 - 700 NTU / FNU / TE/F / FTU
0 - 175 EBC
Regulatory:EPA
Repeatability:Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution:0.0001 NTU / FNU / TE/F / FTU / EBC
Response time:T90 <45 seconds at 100 mL/min
Sample Flow Rate:100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure:6 bar maximum, compared to air at sample temperature range from 2 - 40 °C
Sample Temperature:2 - 60 °C
Signal Average Time:30 - 90 seconds
Storage conditions:-40 - 60 °C
Stray Light:<10 mNTU
System Check:Yes
Units:NTU, FNU, TE/F, FTU, EBC
Verification:Liquid: Stablcal, Formazin (0.1 - 40 NTU)
Dry: Glass Rod at <0.1 NTU
Weight:2.7 kg (5.0 kg with all accessories)
What's included?:TU5300sc Turbidimeter, User Manual, Wall Mount, Vial Replacement Tool, Screw Set, Drying Cartridge, Flow Regulator, Service Bracket

Resources