The perfect environment for superior cell growth.
Secure, reproducible CO? incubation with precise humidity control, intelligent hygiene design and advanced connectivity features.
Product Overview
Range of functions
Calibrate
Three freely selectable values: temperature, humidity, and CO2
Interfaces
USB-C; Ethernet LAN
Timer
Timer - Process time starts immediately when the programme starts; Timer SetpointWAIT - Process time only starts after all setpoint values have been reached; Timer on Event - Process time starts at a defined point in time
Language settings
English (default), German, French, Spanish, Portuguese, Chinese, Arabic, Hindi, Polish, Czech, Hungarian, Russian, Italian, Serbian
Appliance functions
Date and time (time zones, summer/winter time); Display brightness; DistanceView (1–4 freely selectable parameters, activation time adjustable); Network settings (IP, subnet, DHCP); Language and volume settings (buzzer/key tone); Power failure behaviour; Timer functions
6.86" IPS touch display (480 × 1280 pixels) with multi-touch function for graphical display of actual/setpoint values and parameter curves.
O2 (only with T6 option)
Control system
Anaerobic O2 control for targeted oxygen reduction via nitrogen inlet. Zirconium oxide sensor with integrated monitoring and error display. Filtered N? supply for safe process conditions.
Adjustment precision
0,1 %
O2 - Setting range
1 to 20 %
Temperature
Temperature control system
Large-area multi-zone multifunctional all-round heating with additional door, seal and rear wall heating
Setting temperature range
+20 to +50 °C
Working-temperature range
5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting accuracy temperature
0.1 °C
Temperature variation in chamber
at + 37 °C +/- 0.3 K
Temperature fluctuations with time
at 37 °C +/- 0.1 K
Humidity
Humidification system
Passive humidification by means of integrated water tank with condensate return via magnetically guided sealing lip. Humidity limitation by Peltier element. / Active closed-loop humidity control (option K7) with capacitive sensor for precise measurement and complete documentation. Humidification via self-priming micro-dosing pump with hot steam generation, dehumidification via powerful diaphragm pump.
Accuracy
0,5% rh
Setting range
40 to 98 and rh-Off
CO2
Control system
Precise CO2 control with dual-beam NDIR technology, integrated monitoring and pressure-compensated measurement. Filtered CO2 supply for high process reliability.
Temporal fluctuation
± 0,2 %
Setting range
0 to 20 %
Safety
Power failure
Manual mode / Power failure less than 24 hours: Active process is terminated, parameters remain stored. Power failure longer than 24 hours: Operation continues with previously set parameters. Program mode (power failure >1 hour) / Failure less than 1 hour: Programme continues with remaining runtime. Failure longer than 1 hour: Programme is stopped; query for continuation required
Temperature monitoring
Dual Pt1000 sensor, DIN Class A, for mutual monitoring and automatic takeover in the event of a fault
Alarm
visual and acoustic
Standard equipment
Standard option
Door-open detection | V5
Factory calibration certificate III (Option T6)
+37 °C, 5 %CO2, 90 %RH, 10 %O2
Factory calibration certificate II (Option K7)
+37 °C, 5 %CO2, 90 RH
Factory calibration certificate I (Standard)
+37 °C, 5 %CO2
Additional accessories
CO2 connection set, hose with connection coupling and hose clamp; Water tank | B38737; Water tank including connection hose; Stored sterilisation programme: 60 minutes at +180 °C (without removing the sensors)
Door
fully insulated stainless steel door with 2-point locking (compression door lock)
Door
inner glass door with opening (8 mm Ø) to take gas sample
Internals
1 perforated stainless steel shelf/shelves
Stainless steel interior
Dimensions
w(A)x h(B)x d(C): 400 x 425 x 330 mm
Inner cover
37,5 mm
Volume
56 l
Max. number of internals
4
Max. loading of chamber
50 kg
Max. loading per internal
15 kg
Textured stainless steel casing
Dimensions
w(D)x h(E)x d(F): 548 x 628 x 519 mm
Installation
4 adjustable appliance feet
Electrical data
Voltage Electrical load
230 V, 50/60 Hz
approx. 1100 W
Voltage Electrical load
115 V, 50/60 Hz
approx. 1100 W
Ambient conditions
Set Up
The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.