Sumber Aneka Karya Abadi - Your trusted partner for laboratory instrument

Search

POLYMETRON, Polymetron 9611 sc Phosphate Analyser

Prod #9611.KTO.PX.XXX

Product Overview

Spesifikasi :

Accuracy

LR model: ±4 µg/L or ±4% of reading, whichever is greater;
HR model: ±500 µg/L or ±5% of reading, whichever is greater
(Tested with Hach reagents)

Alarm

Four relays output; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Connection: 0.82 to 2.08 mm2 wire (18 to 14 AWG ), 0.82 mm2 stranded (18 AWG) recommended

Certifications

CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000)
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)

Controller Compatibility

SC200, SC1000

Detection limit

 

LR model: 4 µg/L
HR model: 200 µg/L

Enclosure Rating

IP56 / NEMA 4X

Fitting type

Sample line and sample bypass drain: 6 mm (¼-in.)
Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.)

Flow rate

55 - 300 mL/min

Grab sample

Grab Sample In and Grab Sample Out capability

Light source

Class 1M LED (light emitting diode) with a peak wavelength of 810 nm

Manual Languages

English
French
Spanish
B.Portuguese
Chinese
Japanese
Korean
Thai
German
Italian
C.Portuguese
Czech
Danish
Dutch
Polish
Swedish
Finnish
Bulgarian
Hungarian
Romanian
Lithuanian
Russian
Turkish
Slovak
Slovenian
Croatian
Greek
Estonian

Measuring Principle

Colorimetric

Measuring range

 

LR model: 4 - 5000 µg/L as PO4;
HR model: 200 - 50'000 µg/L as PO4

Mounting

Wall, panel or table

Operating Humidity

5 - 95 % non-condensing (indoor use only)

Operating Temperature Range

5 - 45 °C

Output

4 - 20 mA

Number of channels

1, 2, 4; programmable sequence

Pollution Degree

2/II

Power Requirements (Hz)

50/60 Hz

Power Requirements (Voltage)

100 - 240 V AC, 24 V DC

Protection Class

I

Reagent Consumption

2 L of each reagent every 90 days with 15 minute cycle time

Relays

Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum

Repeatability

LR model: ±1%
HR model: ±500 µg/L or ±5% of reading, whichever is greater

Response Time

Typically, 9 minutes at 25°C (77°F); changes with temperature

Response Time

Typically, 9 minutes at 25°C (77°F); changes with temperature

Sample Pressure

0.14 - 6 bar (to Preset Pressure Regulator)

Sample Temperature

5 - 50 °C

Storage Conditions

-20 - 60 °C

Warranty

2 years

Weight

20 kg without reagents and standards, 36.3 kg with reagents

Resources