SPECIFICATIONS :

Accuracy:± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂
± 10% from 5 - 10 mg/L Cl₂
Air purge:Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar maximum
Alerts:Low and high chlorine. Dirty cell. Low and high sample flow.
Certifications:
CE compliant with: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
South Korea KC Certificate
North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003
Controller Compatibility:SC200, SC1000, SC4500. All controllers sold separately
Cycle Time:2.5 minutes
Dimensions (H x W x D):342 mm x 329 mm x 177 mm
Drain:½-inch ID flexible hose
Enclosure Rating:IP66
Inlet:¼-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure:
0.3 - 5.2 bar supplied to Y-strainer
0.1 - 0.3 bar supplied to analyser
Installation Kit:Standpipe Installation Kit
Interferences:Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃.
Light source:LED, measurement at 510 nm, 1 cm light pathlength
Lower Limit of Detection (LOD):0.03 mg/L
Mounting:Wall mount
Operating Humidity:0 - 90% relative humidity, non-condensing
Operating temperature range:5 - 40 °C
Parameter:Chlorine, free; Chlorine, total residual
Power requirements (Voltage):12 VDC, 400 mA maximum (supplied by the controller)
Range:0 -10 mg/L Cl₂
Reagent Consumption:0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
Sample Flow Rate:60 - 200 mL/min through the instrument
Sample requirements:Y-strainer filtration with 40-mesh screen or higher
Sample Temperature:5 - 40 °C
Weight:4.1 kg
What's included?:CL17sc Chlorine Analyser with Standpipe Installation Kit, and User Manual. User provides screws and/or anchors for mounting.
Hach SC Controller and Reagents are not included, they are sold separately.