TU5200 Benchtop Laser Turbidimeter memberikan presisi dan akuratan yang tinggi, dan dengan sensitivitas yang tinggi sehingga meminimalisir variabel pengujian pengujian. TU5200 Benchtop Laser Turbidimeter dapat menghilangkan nilai ketidakpastian karena menggunakan 360º x 90º Detection Technology.. TU5200 Benchtop Laser Turbidimeter mengurangi waktu yang dibutuhkan untuk analisa kekeruhan dengan hasil akurat dan konsisten di laboratorium.

Product Overview

Accuracy:±2 % plus 0.01 NTU from 0 - 40 NTU; 
±10 % of reading from 40 - 1000 NTU based on Formazin primary standard (at 25 °C)
Calibration Method:For Formazin and Stablcal: 
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range

Custom calibration for up to 6-point calibrations

Degrees 0 - 100 mg/L
SDVB 0 - 1000 NTU
Certifications:CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking
Communication:Link2sc
Dimensions (H x W x D):195 mm x 409 mm x 278 mm
Display:7" WVGA (800 x 480) color TFT-LCD with touch screen
Enclosure Construction:PC/ABS Blend Plastic
Light Source:Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Model:TU5200 ISO
Operating Humidity:80 % at 30 °C (non condensing)
Operating Temperature Range:10 - 40 °C
Power Requirements (Hz):50/60 Hz
Power Requirements (Voltage):100 - 240 VAC
Range:EPA:
0 - 700 NTU / FNU / TE/F / FTU
0 - 100 mg/L
0 - 175 EBC

ISO:
0 - 1000 NTU / FNU / TE/F / FTU
0 - 100 mg/L
0 - 250 EBC
Regulatory:DIN EN ISO 7027
Repeatability:<40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
>40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F)
Resolution:0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
Response Time:Signal averaging off: 7 seconds
Signal averaging on: 10 seconds (when averaging time is 5 seconds)
Sample Temperature:4 - 70 °C
Storage Conditions:-30 - 60 °C (-22 - 140 °F)
Stray Light:<10 mNTU
Units:NTU, FNU, TE/F, FTU, EBC;

mg/L if calibrated with Degrees calibration curve
Warranty:1 year
Weight:2.4 kg (5.29 lbs.)