Accuracy : 0.01 - 2 ppb: ± 0.1 ppb
2 - 10,000 ppb: ± 5%
Acidity : < 50 ppm
Ambient Temperature : 5 - 50 °C (41 - 122 °F)
Analog Outputs:6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum
Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) recommended, twisted pair shielded wire
Calibration Method : Automatic with known addition
Manual: 1 or 2 points
Dimensions : 681 mm x 452 mm x 254 mm (H x W x D)
Display : Colored 5.7” LCD
Electrode Type : Sodium ISE (ion specific electrode) electrode and reference electrode with KCl electrolyte
Fuse : Input power: T 1.6 A, 250 VAC
Relays: T 5.0 A, 250 V
Inlet : Sample line and sample bypass drain: 6 mm O.D. push-to-connect fitting for plastic tubing
Chemical and case drains: 7/16 inch I.D. slip-on fitting for soft plastic tubing
Interference Phosphate 10 ppm : < 0.1 ppb
Load of Analog Outputs : 600 Ohm
Lower Limit of Detection (LOD) : 0.01 ppb
Maintenance Interva l: Every 90 days: refill electrolyte, reactivation, conditioning, and calibration solution
Material : Polyol case, PC door, PC hinges and latches, 304/316 SST hardware
Max. Concentration of Suspended Solids in Sample : < 2 NTU, no oil, no grease
For boiler sample type install approx. 100 µm filter
Mounting : Panel mount
Number of analog outputs : 6
Number of Channels : 1
Number of Relays: 6
Options:Analyzer only
Parameter:Sodium
pH Range:6 - 10 pH
Pollution Degree:2
Power Requirements (Hz):50/60 Hz
Power Requirements (Voltage):100 - 240 VAC
Protection Rating:IP65, PCBA housing
Range:0.01 ppb - 10,000 ppb
Relative Humidity:10 - 80%, non-condensing
Relay Output:6; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Connection: 1.0 - 1.29 mm² (18 - 16 AWG) wire; 1.0 mm² (18 AWG) stranded recommended, 5 - 8 mm O.D. cable
Repeatability:< 0.02 ppb or 1.5% reading (whichever is greater) within ±10 °C (±50 °F) variation
Response Time : From 0.1 ppb to 10 ppb: T90 ≤ 3 minutes, T95 ≤ 4 minutes
From < 1 ppb to 100 ppb: T90 < 2 minutes, T95 < 3 minutes (about 150 s)
Sample conditioner : Di-isopropylamine (DIPA) (1 L/100 days) at 25 °C for a sample pH target of 10.5
Sample Flow Rate : 100 - 150 mL/min (6 - 9 L/h)
Sample Pressure : 0.2 - 6 bar (3 - 87 psi)
Sample Temperature : 5 - 45 °C (41 - 113 °F)
Storage Conditions : -20 - 60 °C (2 -140 °F)
Weight : 14 kg (30.7 lb) with empty bottle